Fast Measurement of Electrochemical Impedance Spectroscopy Based on White Noise
Received:April 09, 2020   Revised:April 15, 2020   Accepted:April 18, 2020      Published Online:January 29, 2021
View Full Text  View/Add Comment  Download reader
DOI:
KeyWord:Electrochemical impedance spectroscopy; Wavelet transform; White noise; Simulink
           
AuthorInstitution
Zhang Xu Kunming University of Science and Technology
YANG Chang-jiang Faculty of Metallurgy and Energy Engineering,Kunming University of Science and Technology
ZHAO Lv-xing Faculty of Metallurgy and Energy Engineering,Kunming University of Science and Technology
GAO Bo-bin Faculty of Metallurgy and Energy Engineering,Kunming University of Science and Technology
Hits: 1832
Download times: 855
Abstract:
      Commercial instruments of electrochemical impedance spectroscopy (EIS) are usually based on Frequency Response Analyzer with sine signal. A fast measurement of electrochemical impedance spectroscopy was provided through white noise signal and wavelet transform and simulation experiments was carried out with the software of Simulink. The effect of the center frequency, sampling rates and time were optimized in terms of the equivalent circuit of (RC)(RC). With impedance range from 0.05Hz to 20000 Hz, the measuring relative error of elements of the equivalent circuit were less than 0.5% and the elapsed time was 85s. At the equivalent circuits of both proton exchange membrane fuel cell and pitting corrosion of passivated metal, the approach could be well applied with the fitting error less than 0.2%. EIS can be obtained fast and directly acquired from the white noise signal of the electrochemical process to achieve an in-site and real-time monitoring at low cost.
Close